SITUATE
Received Nov 5, 2024 · Event occurred Oct 16, 2024
Report 1717344-2024-02051 · MDR key 20612937
Device
Generic name
Gauze/sponge, Internal, X-ray Detectable
Manufacturer
Covidien Mfg Dc BoulderModel number
G0804-12P02CN-1Catalog number
G0804-12P02CN-1
Lot number
221208KF
Product problems
- Material Fragmentation
- Material Fragmentation
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
Description of Event or Problem
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE RAYTECS INCLUDED IN THE HEAD AND NECK PACKS HAD BEEN INCONSISTENT IN QUALITY. MORE SPECIFICALLY, THE BLUE COATED THREAD THAT WAS SPIRALED THROUGHOUT THE SPONGE APPEARED TO BE CHIPPED, CAUSING THE MATERIAL TO SEPARATE FROM THE SPONGE AND ENTER THE OPERATIVE SPACE. THERE WAS NO PATIENT INJURY.
Additional Manufacturer Narrative
D10 CONCOMITANT PRODUCTS: G0804-12P02CN-1 - GAUZE; XRAY; RFD; 8X4; 12-PLY; NSTRL; LF, LOT# 221208KF. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
Description of Event or Problem
ACCORDING TO THE REPORTER, THE RAYTECS INCLUDED IN THE HEAD AND NECK PACKS HAVE BEEN INCONSISTENT IN QUALITY. MORE SPECIFICALLY, THE BLUE COATED THREAD THAT IS SPIRALED THROUGHOUT THE SPONGE APPEARS TO BE CHIPPED, CAUSING THE MATERIAL TO SEPARATE FROM THE SPONGE AND ENTER THE OPERATIVE SPACE. THERE WAS NO PATIENT INJURY.
Additional Manufacturer Narrative
ADDITIONAL INFORMATION: G3, H3, H6 H3. EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS NOT RETURNED, BUT A PHOTO WAS AVAILABLE FOR EVALUATION. VISUAL INSPECTION NOTED THE THREADS OF THE BLUE RAYTEC LINE SEEM TO HAVE MOVED AWAY FROM THE MAIN BLUE THREADS. IT WAS REPORTED THAT THE BLUE COATED THREAD THAT IS SPIRALED THROUGHOUT THE SPONGE APPEARS TO BE CHIPPED. THE REPORTED ISSUE WAS CONFIRMED. THE MOST LIKELY CAUSE COULD NOT BE ESTABLISHED FROM THE INFORMATION AVAILABLE. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.