NRFIT CONNECTOR SET

Received Nov 1, 2024 · Event occurred Oct 1, 2024

Report 3012307300-2024-12562 · MDR key 20589767

Device

Generic name

Administrations Sets With Neuraxial Connectors

Catalog number

21-7605-24

Lot number

4461386

Product problems

  • Obstruction of Flow
  • Obstruction of Flow

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT, WHEN THE PCEA IS CONNECTED, AN ALARM SOUNDS, WHICH IS THOUGHT TO BE DUE TO A BLOCKAGE. AFTER REPLACING IT WITH ANOTHER EXTENSION TUBE, THERE ARE NO PROBLEMS. THIS OCCURRED DURING PRIMING AT THE FACILITY. THERE WAS NO PATIENT INVOLVEMENT.

Additional Manufacturer Narrative

E1: (B)(6). H3: INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative

ONE DEVICE WAS RETURNED FOR ANALYSIS. AS RECEIVED, NO DAMAGES OR ANOMALIES WERE OBSERVED. NO ERRORS OR ALARMS WERE OBSERVED, AND NO LEAKAGE WAS NOTED DURING TESTING AND ISSUE WAS NOT REPLICATED. THE COMPLAINT OF ALARMING WHEN CONNECTED COULD NOT BE REPLICATED OR CONFIRMED.