inforMED
InjuryJDP

4.5MM VA-LCP CURVED CONDYLAR PLATE/16HOLE/336MM/RT-STER

Received Oct 30, 2024 · Event occurred Oct 12, 2024

Report 8030965-2024-13398 · MDR key 20566655

Device

Generic name

Condylar Plate Fixation Implant

Manufacturer

Synthes Gmbh

Catalog number

02.124.416S

Lot number

7038P92

Product problems

  • Break
  • Break

Patient

51 YR · Female

  • Bone Fracture(s)
  • Bone Fracture(s)

Narrative

Additional Manufacturer Narrative

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H3, H4, H6: THE PRODUCT WAS NOT RETURNED TO MEDTECH ORTHOPAEDICS, HOWEVER PHOTOS WERE PROVIDED FOR REVIEW. VISUAL ANALYSIS OF THE X-RAYS SERIES REVEALED THAT VA-LCP CONDYLAR PLATE 4.5/5.0 R 16HO L33 WAS ABLE TO IDENTIFY BREAKAGE ACROSS THE SHAFT. ALL FRAGMENTS WERE OBSERVED IN THE VISUAL EVIDENCE PROVIDED. WITH THE INFORMATION PROVIDED IS NOT POSSIBLE TO DETERMINE A POTENTIAL CAUSE AT THIS MOMENT. THE MODE OF FAILURE OF THE DEVICE IS MULTI-FACTORIAL AND CONSIDERATION MUST BE GIVEN TO ALL OTHER POTENTIAL INFLUENCES SUCH AS SURGICAL PROCESS, PATIENT VARIABLES (E.G. ACTIVITY LEVEL AND USE), ANATOMICAL CONSIDERATIONS AND PATIENT CHANGES OVER TIME. ANY CONCLUSIONS FROM THE INVESTIGATIONAL INPUT PROVIDED HAVE TO BE PLACED INTO CONTEXT WITH ALL OTHER RELEVANT FACTORS. SINCE THE DEVICE WAS NOT RETURNED, A DIMENSIONAL INSPECTION CANNOT BE PERFORMED. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE VA-LCP CONDYLAR PLATE 4.5/5.0 R 16HO L33 WOULD CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION. AS PART OF MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY REVIEW : PART NUMBER: 02.124.416S, LOT NUMBER: 7038P92, MANUFACTURING SITE: MEZZOVICO, RELEASE TO WAREHOUSE DATE: 28 AUG 2023, EXPIRATION DATE: 01 AUG 2033. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR STERILE FINISHED LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Description of Event or Problem

IT WAS REPORTED THAT FIVE MONTHS POSTOPERATIVELY, PATIENT WAS WALKING WHEN THEIR LEG GAVE WAY; THE PLATE WAS FOUND TO HAVE BROKEN IN SITU. PATIENT UNDERWENT REVISION SURGERY ON (B)(6) 2024, USING TITANIUM LISS.

Additional Manufacturer Narrative

THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.