inforMED
MalfunctionFAM

PENTAX

Received Oct 28, 2024 · Event occurred Oct 7, 2024

Report 9610877-2024-57838 · MDR key 20542162

Device

Generic name

Video Sigmoidoscope 4.2c 12.8t 700l Fwj

Model number

ES-3870K

Product problems

  • Erratic or Intermittent Display
  • Erratic or Intermittent Display

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD MODULE WITH DRIVE PCB DISTORTED IMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD MODULE WITH DRIVE PCB. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.