ELECSYS T4
Received Oct 25, 2024 · Event occurred Oct 1, 2024
Report 1823260-2024-03082 · MDR key 20533698
Device
Generic name
Radioimmunoassay, Total Thyroxine
Manufacturer
Roche DiagnosticsModel number
T4Catalog number
09007784190
Lot number
790382
Product problems
- High Test Results
- Non Reproducible Results
- High Test Results
- Non Reproducible Results
Patient
50 YR · Female
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE SERIAL NUMBER OF THE COBAS E 801 ANALYTICAL UNIT IS (B)(6). THE INVESTIGATION IS ONGOING.
Description of Event or Problem
THE INITIAL REPORTER RECEIVED QUESTIONABLE ELECSYS T4 (T4) RESULT FROM ONE PATIENT SAMPLE TESTED ON THE COBAS E 801 ANALYTICAL UNIT. THE INITIAL RESULT WAS 16.9 MG/DL. THIS RESULT WAS DEEMED TOO HIGH FOR THE "PATIENT'S CLINIC". THE FIRST REPEAT RESULT WAS 5.78 MG/DL. THE SECOND REPEAT RESULT WAS 5.46 MG/DL.
Additional Manufacturer Narrative
THE INVESTIGATION NOTED THAT THE INITIAL AND REPEAT RESULTS WERE FROM DIFFERENT REAGENT PACKS. THE FIRST REAGENT PACK WHICH GAVE THE INITIAL QUESTIONABLE RESULT HAD FILM/FOAM/BUBBLES ON THE SURFACE, CAUSING THE PIPETTING OF AN INSUFFICIENT REAGENT VOLUME. THE INVESTIGATION REVIEWED THE CALIBRATION DATA FROM THE SECOND REAGENT PACK; THE RESULTS WERE WITHIN SPECIFICATIONS BUT ON THE LOW SIDE. THE INVESTIGATION REVIEWED THE QC DATA. THE QC DATA OF THE FIRST REAGENT PACK HAD AN ERROR ALARM. THE QC DATA OF THE SECOND REAGENT PACK WERE WITHIN SPECIFICATIONS. THE FIELD SERVICE ENGINEER PERFORMED PREVENTIVE MAINTENANCE ON THE ANALYZER AND REPLACED THE MEASURING CELLS. HE PERFORMED AN INSTRUMENT CHECK WITH SUCCESSFUL RESULTS. A GENERAL REAGENT ISSUE CAN BE EXCLUDED BASED ON THE INFORMATION PROVIDED. THE INVESTIGATION DETERMINED THE EVENT WAS CONSISTENT WITH FILM/FOAM/BUBBLES ON THE REAGENT SURFACE PRODUCT LABELING STATES "AVOID FOAM FORMATION IN ALL REAGENTS AND SAMPLE TYPES (SPECIMENS, CALIBRATORS AND CONTROLS)."