inforMED
InjuryHRS

2.4/2.7 VA-LCP 2-C DRP STD 6H HD/10H SHAFT/145MM/RT-STER

Received Oct 24, 2024

Report 2939274-2024-00005 · MDR key 20523009

Device

Generic name

Plate, Fixation, Bone

Catalog number

02.111.680S

Product problems

  • Adverse Event Without Identified Device or Use Problem
  • Adverse Event Without Identified Device or Use Problem

Patient

NA · Unknown

  • Pain
  • Skin Inflammation/ Irritation
  • Pain
  • Skin Inflammation/ Irritation

Narrative

Additional Manufacturer Narrative

THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. ADDITIONAL NARRATIVE: D4: UDI: AS THE LOT NUMBER FOR THE DEVICE INVOLVED IN THE EVENT WAS NOT PROVIDED, THE FULL UDI IS CURRENTLY NOT AVAILABLE.

Description of Event or Problem

THIS COMPLAINT IS FROM A DATA USE AGREEMENT (DUA). THE DUA SUMMARIZES 55 PATIENTS WHO UNDERWENT WRIST SURGERY BETWEEN 2008 AND 2023 UTILIZING DEPUY SYNTHES LCP DIA-META VOLAR DISTAL RADIUS PLATES (N=20), DEPUY SYNTHES 2.4 MM VA-LCP DORSAL DISTAL RADIUS PLATES (N=20), AND DEPUY SYNTHES 2.4/2.7 MM VA-LCP TWO-COLUMN VOLAR DISTAL RADIUS PLATE (N=15), EXTRA-LONG AT CLEVELAND CLINIC FOUNDATION (CCF). PATIENT IS A 71-YEAR-OLD FEMALE. REASON FOR IMPLANTATION WAS FRACTURE REPAIR. POST OP COMPLICATION INCLUDE RETAINED HARDWARE RIGHT FOREARM, PAINFUL HARDWARE/IRRITATION & PROMINENT CREPITANCE. TREATMENT/INTERVENTION INCLUDE HARDWARE REMOVAL, RIGHT FOREARM. IMPLANT(S) INVOLVED: PLATE X1 ,OF 2.4 & 2.7 SCREWS X 11.

Additional Manufacturer Narrative

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H6 COMPONENT CODES: MOST RELEVANT COMPONENT CODE IS G07002 (APPROPRIATE TERM/CODE NOT AVAILABLE) TO CAPTURE NO FINDINGS AVAILABLE DUE TO NO PRODUCT RETURNED. INVESTIGATION SUMMARY: PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.