inforMED
MalfunctionFPK

MEDEX STOPCOCK EXTENSION

Received Oct 22, 2024 · Event occurred Oct 2, 2024

Report 9616567-2024-00141 · MDR key 20507898

Device

Generic name

Tubing, Fluid Delivery

Catalog number

MX43860

Lot number

4457074

Product problems

  • Break
  • Break

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE.

Description of Event or Problem

IT WAS REPORTED THAT THE EXTENSION TUBING BROKE OFF AT THE STOPCOCK CONNECTION. THERE WAS PATIENT INVOLVEMENT, AND NO HARM REPORTED.

Additional Manufacturer Narrative

D9: DATE RETURNED TO MFG.: 10/9/2024. H3 AND H6. CODES: UPDATED. ONE DEVICE WAS RECEIVED FOR INVESTIGATION. THE COMPLAINT WAS CONFIRMED. THERE WAS SEPARATION OF THE SOLVENT BOND OF STOPCOCK AND TUBE. ADDITIONALLY, IT WAS OBSERVED SOLVENT SIGNATURE WAS PRESENT ON THE TUBE. THE CAUSE WAS MANUFACTURING. A DEVICE HISTORY RECORD (DHR) REVIEW REPORTED NO DISCREPANCIES OR NON-CONFORMANCES DURING THE MANUFACTURING OF THE REPORTED LOT NUMBER. AWARENESS WAS MADE TO OPERATION PERSONNEL WHO PERFORM THE ASSEMBLY PROCESS, AND THE ISSUE WILL BE MONITORED FOR THRESHOLD OR ESCALATION.