inforMED
MalfunctionGXZ

APS¿ (AUTOMATIC PERIODIC STIMULATION) ELECTRODE

Received Oct 21, 2024 · Event occurred Sep 18, 2024

Report 9612501-2024-02679 · MDR key 20491363

Device

Generic name

Electrode, Needle

Model number

8228052

Catalog number

8228052

Lot number

0228844531

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings
  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

53 YR · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

H3: PRODUCT ANALYSIS FOUND THAT THE C-CLIP CONTACT WAS NOT EXPOSED/FLUSH WITH THE INSIDE DIAMETER OF THE CLIP BODY UPON RETURN. THE CONTACT WAS RECESSED INSIDE THE CLIP BODY BY 0.06 INCHES (FROM THE CLIP OPENING TO THE DISTAL TIP OF THE CONTACT). THE RESISTANCE SHALL BE LESS THAN 25 OHMS END TO END, AND THE ACTUAL MEASUREMENT WAS 11.4 OHMS WHICH WAS IN SPECIFICATION. THE COMPLAINT WAS CONFIRMED, DUE TO AN OUT OF SPECIFICATION CONDITION. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative

B5: EVENT DESCRIPTION HAS BEEN UPDATED. H3: ANNEX F CODE HAS BEEN UPDATED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem

ADDITIONAL INFORMATION RECEIVED STATING THE METAL PIN IN FRONT OF THE FIRST APS WAS MISSING, THERE WAS NO SIGNAL. THE OTHER THREE APS DEVICES WERE FOUND DEFECTIVE UPON THE UNBOXING, SO THEY HAD NOT BEEN USED. PROCEDURE WAS COMPLETED WITHOUT ANY PROBLEM. THERE WAS NO PATIENT IMPACT.

Description of Event or Problem

IT WAS REPORTED THAT APS ELECTRODE WAS NOT STIMULATING. THERE WAS NO KNOWN PATIENT IMPACT.

Additional Manufacturer Narrative

MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.