inforMED
MalfunctionGDG

VERITY HOOK, NERVE, TITANIUM HANDLE, 3 MM BALL TIP, 7 1/4 INCH SHAFT

Received Oct 14, 2024 · Event occurred Sep 13, 2024

Report 3007208013-2024-00053 · MDR key 20441973

Device

Generic name

Surgical Hook

Model number

80-1531

Lot number

UNKNOWN

Product problems

  • Break
  • Break

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE DEVICE WAS NOT RETURNED FOR EVALUATION. THE LOT NUMBER OF THE DEVICE AND PICTURES OF THE PROBLEMATIC DEVICE WERE NOT PROVIDED. BASED ON THIS, WE WERE VERY LIMITED IN OUR ABILITY TO INVESTIGATE. THE COMPLAINT COULD NOT BE CONFIRMED, AND THE ROOT CAUSE CANNOT BE DETERMINED. IF ADDITIONAL RELEVANT INFORMATION IS IDENTIFIED, IT WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative

SOME PICTURES HAVE BEEN PROVIDED. PRODUCT IS BEING SENT BACK FOR EVALUATION. WE DO NOT HAVE LOT INFORMATION YET BUT SHOULD BE ABLE TO GET THAT FROM THE DEVICE ONCE IT IS RETURNED. ONCE THE INVESTIGATION IS COMPLETE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem

COMPLAINANT ALLEGES "PART 80-1531 VERITY NERVE HOOK HAD THE BALL TIP BROKEN OFF INTO THE PATIENT. USED IN ACDF CASE. REQUIRED RETRIEVAL OF THE TIP, AND AN X-RAY TO CONFIRM ALL PIECES WERE RETRIEVED. NO HARM TO THE PATIENT."