inforMED
MalfunctionQOF

COBAS® SARS-COV-2 & INFLUENZA A/B

Received Oct 11, 2024 · Event occurred Sep 12, 2024

Report 2243471-2024-02960 · MDR key 20428577

Device

Generic name

Multi-target Resp Specimen Nat Incl. Sars-cov-2 & Other Microbial Agents

Catalog number

09211101190

Lot number

40422A

Product problems

  • False Positive Result
  • False Positive Result

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THERE WAS AN ALLEGATION OF QUESTIONABLE SARS-COV-2 RESULTS WITH 6 PATIENT SAMPLES USING THE COBAS SARS-COV-2 & INFLUENZA A/B ASSAY FOR USE ON THE COBAS LIAT ANALYZER. THE ALLEGED SAMPLES INITIALLY GENERATED A POSITIVE RESULT FOR SARS-COV-2. THE SAME SAMPLES WERE REPEATED USING THE SAME OR A DIFFERENT LIAT ANALYZER GENERATING A NEGATIVE SARS-COV-2 RESULT. THE NEGATIVE RESULTS WERE DEEMED CORRECT.

Additional Manufacturer Narrative

THE SERIAL NUMBERS OF THE COBAS LIAT ANALYZER WERE (B)(6). INVESTIGATION IS ONGOING.

Additional Manufacturer Narrative

ROCHE HAS RECEIVED COMPLAINTS FROM CUSTOMERS USING COBAS® SARS-COV-2 & INFLUENZA A/B TEST FOR USE ON THE COBAS® LIAT SYSTEM, LOT 40422A. CUSTOMERS ARE ALLEGING AN INCREASED NUMBER OF INITIAL POSITIVE SARS-COV-2 RESULTS THAT WERE NEGATIVE UPON RETESTS ON THE SAME PLATFORM AND/OR ON OTHER PLATFORMS. BASED ON AN INVESTIGATION OF RECENT CASES, THERE IS A LOW OCCURRENCE RATE OF THIS ISSUE WITH COBAS ® SARS-COV-2 & INFLUENZA A/B KIT LOT 40422A. AT THIS OCCURRENCE RATE, THE GENERATION OF FALSE POSITIVE RESULTS IS WITHIN THE US-IVD PRODUCT CLINICAL PERFORMANCE CLAIMS FOR NEGATIVE PERCENT AGREEMENT AT 99.4% (95% CI: 98.3 - 99.8) AND THE OVERALL PRODUCT BENEFITS FOR THE RAPID AND ACCURATE DETECTION OF SARS-COV-2 AND INFLUENZA A/B AS AN AID IN THE DIFFERENTIAL DIAGNOSIS OF THESE INFECTIONS ARE CONSIDERED TO OUTWEIGH THIS RISK WHEN USED ACCORDING TO ITS INTENDED USE. TAKING THE LOW RATE OF FALSE POSITIVE OCCURRENCE AND THE LOW PROBABILITY OF SUBSEQUENT PATIENT SAFETY IMPACTS TOGETHER, THE FALSE POSITIVES ARE UNLIKELY TO CAUSE ADVERSE TRANSIENT OR SEVERE ADVERSE EVENTS. ADDITIONALLY, NO HARM WAS INDICATED IN THE CURRENT CASE. THE PRELIMINARY INVESTIGATION INTO THE COMPLAINT LOT HAS REPRODUCED THE ISSUE AND FINDINGS SUGGEST REAGENT CONTAMINATION AS A POTENTIAL CAUSE. CUSTOMERS HAVE BEEN INSTRUCTED TO DISCONTINUE THE USE OF AND DISCARD ANY REMAINING INVENTORY OF THE COBAS® SARS-COV-2 & INFLUENZA A/B NUCLEIC ACID TEST FOR USE ON THE COBAS ® LIAT SYSTEM, LOT 40422A, AND ROCHE WILL RECORD THE ACTIONS PER 21 CFR 806.