COBAS® PCR MEDIA UNI SWAB SAMPLE KIT
Received Oct 10, 2024 · Event occurred Sep 2, 2024
Report 2243471-2024-02945 · MDR key 20418158
Device
Generic name
Specimen Collection Kit
Manufacturer
Roche Molecular Systems, Inc. - BranchburgCatalog number
07958030190
Lot number
77698501
Product problems
- Break
- Break
Patient
NA · Unknown
- Unspecified Gastrointestinal Problem
- Unspecified Gastrointestinal Problem
Narrative
Additional Manufacturer Narrative
THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. AN OFF-LABEL SAMPLE COLLECTION WAS PERFORMED. PER THE OPERATOR'S MANUAL, OROPHARYNGEAL SWAB COLLECTIONS ARE INTENDED TO BE PERFORMED BY A CLINICIAN. IT IS POSSIBLE THAT THE PATIENT ACCIDENTALLY BROKE THE SWAB AT THE SCORELINE DURING COLLECTION.
Description of Event or Problem
A CUSTOMER ALLEGED THAT A SWAB, WHICH IS PACKAGED IN THE COBAS® PCR MEDIA UNI SWAB SAMPLE KIT, BROKE DURING SAMPLE COLLECTION. THE PATIENT WAS ALLEGEDLY COLLECTING AN OROPHARYNGEAL SWAB WHEN THE SWAB BROKE DURING SELF-COLLECTION. IT IS UNCLEAR AT WHAT POINT IN THE COLLECTION PROCESS THE SWAB BROKE AND IF ANY TREATMENT WAS ADMINISTERED. NO FURTHER HARM WAS REPORTED.