PATCH
Received Apr 5, 2011 · Event occurred Jan 4, 2011
Report 2124215-2011-00424 · MDR key 2039680
Device
Product problems
- High impedance
- Material Integrity Problem
Patient
74 YR
- No Consequences Or Impact To Patient
Narrative
Description of Event or Problem
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DETECTED OUT OF RANGE HIGH SHOCK IMPEDANCES THROUGH THESE DEFIBRILLATION PATCHES. THERE WAS REPORT THAT DURING A CHANGE-OUT PROCEDURE THE PHYSICIAN HAD REPAIRED INSULATION ON ONE OF THE PATCHES. TECHNICAL SERVICES DISCUSSED OPTIONS AND TROUBLESHOOTING. NO ADVERSE PATIENT EFFECTS WERE REPORTED. RESOLUTION HAS BEEN REQUESTED FROM THE FIELD REPRESENTATIVE. IT WAS LATER REPORTED THAT DEFIBRILLATION THRESHOLDS WERE TESTED AND THE SHOCK VECTOR WAS CHANGED WHICH RESOLVED THE IMPEDANCES.
Additional Manufacturer Narrative
INFORMATION SUGGESTS THESE PRODUCTS REMAIN IN-SERVICE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS EVENT WILL BE UPDATED.