inforMED
MalfunctionITQ

DONJOY

Received Oct 4, 2024 · Event occurred Sep 18, 2024

Report 9616086-2024-00085 · MDR key 20378303

Device

Generic name

X-rom Post-op Knee Brace

Manufacturer

Djo Llc

Model number

11-2181-9

Lot number

UNKNOWN

Product problems

  • Mechanical Problem
  • Unintended System Motion
  • No Apparent Adverse Event
  • Mechanical Problem
  • Unintended System Motion
  • No Apparent Adverse Event

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

IT WAS REPORTED THAT ''BRACE WILL NOT ENGAGE (LOCK) INTO PRESET POSITIONS (RED BUTTON); JUST OUT OF THE BAG, BAG WAS DISCARDED. ONE DAY OF USE''. THE BRACE HAS NOT BEEN RETURNED FOR EVALUATION. THE ROOT CAUSE OF THE COMPLAINT IS DAMAGED COMPONENT BASED ON OTHER DEVICES THAT HAVE BEEN EVALUATED. THE DEVICES LOCK BUTTON FAILED DURING MANUFACTURER TESTING CONFIRMED TO BE THE SAME MALFUNCTION FOUND IN REPORT NUMBER 9616086-2023-00007. ADDITIONAL REPORTING ON THIS EVENT WILL BE PROVIDED AS A SUPPLEMENTAL REPORT TO THIS DOCUMENT IF IT BECOMES AVAILABLE.

Description of Event or Problem

IT WAS REPORTED THAT ''BRACE WILL NOT ENGAGE (LOCK) INTO PRESET POSITIONS (RED BUTTON); JUST OUT OF THE BAG, BAG WAS DISCARDED. ONE DAY OF USE''.