inforMED
MalfunctionGAD

FASTSYSTEM STERILE FIELD POST

Received Oct 4, 2024

Report 3004608878-2024-00118 · MDR key 20370720

Device

Generic name

Surgical Retractor

Catalog number

4080

Lot number

6212276

Product problems

  • Mechanics Altered
  • Mechanics Altered

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE FASTSYSTEM STERILE FIELD POST (4080) WAS RETURNED FOR EVALUATION: THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED, AND NO ANOMALIES RELATED TO THE REPORTED FAILURE WERE OBSERVED. FAILURE ANALYSIS: EVALUATION FOUND THAT THE CLAMP JAW OF THE STERILE FIELD POST WAS UNSTABLE WHEN MOUNTED ON THE TEST RAIL. THE BOLT HAD BECOME FLAT AT THE BOTTOM AND MUST BE REPLACED ALONG WITH THE WASHERS. ROOT CAUSE ANALYSIS: THE COMPLAINT REPORTED BY THE CUSTOMER WAS CONFIRMED AS VALID. BASED ON THE EVALUATION BY INTEGRA SERVICE AND REPAIR, THE ROOT CAUSE WAS MOST LIKELY USE-ERROR (CLOSING THE CLAMP WITH NOTHING INSIDE). NO MANUFACTURING, WORKMANSHIP, OR MATERIAL DEFICIENCY WAS IDENTIFIED FOR CORRECTION AS A RESULT OF THIS COMPLAINT INVESTIGATION.

Description of Event or Problem

IT WAS REPORTED THAT THE FASTSYSTEM STERILE FIELD POST (4080) DID NOT STAY TIGHTENED WHEN ATTACHED TO THE TABLE DURING SURGERY. A DELAY OF AT LEAST 30 MINUTES WAS REPORTED AS A RESULT OF LOOKING FOR ANOTHER RETRACTOR USED TO COMPLETE THE SURGERY. THE WAS NO PATIENT INJURY OR DEATH REPORTED.