inforMED
MalfunctionGCL

PENTAX

Received Oct 1, 2024 · Event occurred Sep 23, 2024

Report 9610877-2024-57195 · MDR key 20341575

Device

Generic name

Video Esophagoscope

Model number

EE17-J10

Product problems

  • Obstruction of Flow
  • Obstruction of Flow

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

G4: THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EE-1580K IS AVAILABLE IN THE USA WITH A 510K NUMBER K961564. D4 UDI: "NOT DISTRIBUTED IN USA", BECAUSE PRODUCTS ARE NOT DISTRIBUTED IN USA PRODUCTS, BUT SIMILAR DEVICE PRODUCT ARE LISTED IN USA. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE AIR/WATER SOCKET CLOGGED. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE INADEQUATE/INSUFFICIENT REPROCESSING AT THE FACILITY ON THE AIR/WATER SOCKET. IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE OBJECTIVE LENS LEAK; HOWEVER, THIS DEFECT IS NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. THIS DEFECT OCCURRED NOT DURING PROCEDURE. BASED ON THE TECHNICAL REPORT ""HR-RPT-0630(AIR/WATER & JET WATER CHANNELS)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem

THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. AIR/WATER TUBE CLOGGED.