LOOP CUTTER
Received Sep 23, 2024 · Event occurred Sep 4, 2024
Report 9614641-2024-01878 · MDR key 20285588
Device
Product problems
- Difficult to Remove
- Difficult to Remove
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE EVALUATION OF THE EVENT IS ONGOING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem
IT WAS REPORTED DURING AN UNSPECIFIED THERAPEUTIC PROCEDURE WHILE CUTTING THE LOOP OF THE SINGLE USE LIGATING DEVICE WITH THE LOOP CUTTER, THE LOOP GOT CAUGHT IN THE CUTTER BLADE AND COULD NOT BE REMOVED. THE USER ATTEMPTED TO OPERATE THE HANDLE PART WHILE KEEPING THE ENDOSCOPE AS STRAIGHT AS POSSIBLE BUT THERE WAS NO IMPROVEMENT. A SPARE LOOP CUTTER WAS USED TO CUT THE DEVICES AND REMOVE THEM FROM THE PATIENT. THE PROCEDURE WAS THEN COMPLETED. THERE WAS NO REPORTED PATIENT IMPACT.
Additional Manufacturer Narrative
THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION AND DEVICE EVALUATION. ADDITIONALLY, TO PROVIDE CORRECTION TO THE INITIAL WITH INFORMATION INADVERTENTLY LEFT OUT (D9 AND G2) AND TO PROVIDE AN UPDATE TO FIELDS (D4, H3, AND H4). THE DEVICE WAS EVALUATED BY OLYMPUS. THE INSERTION SECTION WAS SEVERED APPROXIMATELY 222 CM FROM THE TIP, AND THE LOOPS TRAPPED IN THE CUTTER STORAGE AREA MEASURED AROUND 2.5 MM IN LENGTH. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THAT THE REPORTED EVENT OCCURRED DUE TO THE USER TRIED TO CUT THE LOOP WITHOUT IT RESTING ON BOTH SIDES OF THE LOOP HOLDER, THE LOOP GOT STUCK IN THE CUTTER HOUSING, MAKING IT IMPOSSIBLE TO CUT THE LOOP AND REMOVE IT FROM THE LOOP. HOWEVER, THE SPECIFIC ROOT CAUSE OF THE REPORTED EVENT COULD NOT BE DETERMINED. THE EVENT CAN BE PREVENTED BY FOLLOWING THE INSTRUCTIONS FOR USE WHICH STATE: "BEFORE EACH CASE, PREPARE AND INSPECT THE INSTRUMENT AS INSTRUCTED BELOW. INSPECT OTHER EQUIPMENT TO BE USED WITH THE INSTRUMENT AS INSTRUCTED IN THEIR RESPECTIVE INSTRUCTION MANUALS. SHOULD THE SLIGHTEST IRREGULARITY BE SUSPECTED, DO NOT USE THE INSTRUMENT; CONTACT OLYMPUS. DAMAGE OR IRREGULARITY MAY COMPROMISE PATIENT OR USER SAFETY, SUCH AS PUNCTURES, HEMORRHAGES OR MUCOUS MEMBRANE DAMAGE AND MAY RESULT IN MORE SEVERE EQUIPMENT DAMAGE." "ALWAYS HAVE A SPARE INSTRUMENT AVAILABLE." "DO NOT TRY TO CUT THE LOOP THAT IS NOT POSITIONED ON BOTH EDGES OF THE LOOP HANGER AS PLUMB AS POSSIBLE FOR THE BLADE. IT MAY MAKE CUTTING THE LOOP IMPOSSIBLE, OR RESULT IN THE LOOP GETTING CAUGHT IN THE DISTAL END OF THE INSTRUMENT, WHICH COULD MAKE IT DIFFICULT OR IMPOSSIBLE TO REMOVE FROM THE PATIENT. IN THIS CASE, USE PLIERS TO CUT THE INSERTION PORTION OF THE INSTRUMENT WHERE IT EXTENDS FROM THE BIOPSY VALVE OF THE ENDOSCOPE. REMOVE THE ENDOSCOPE FROM THE BODY, THEN REINSERT THE ENDOSCOPE AND CUT THE LOOP WITH A SPARE LOOP CUTTER." "DO NOT CUT THE LOOP UNLESS YOU HAVE A CLEAR ENDOSCOPIC FIELD OF VIEW. THIS COULD CAUSE PATIENT INJURY, SUCH AS PUNCTURES, HEMORRHAGES OR MUCOUS MEMBRANE DAMAGE. IT MAY ALSO DAMAGE THE ENDOSCOPE AND/OR INSTRUMENT." "IF IT IS DIFFICULT TO CUT TWO LOOPS SIMULTANEOUSLY, CUT THEM ONE AT A TIME. FORCIBLE CUTTING MAY DAMAGE THE LOOP CUTTER." OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.