inforMED
MalfunctionITQ

DONJOY

Received Sep 9, 2024 · Event occurred Aug 21, 2024

Report 9616086-2024-00075 · MDR key 20177223

Device

Generic name

X-rom Post-op Knee Brace

Manufacturer

Djo Llc

Model number

11-2181-9

Lot number

UNKNOWN

Product problems

  • Mechanical Problem
  • Unintended System Motion
  • No Apparent Adverse Event
  • Mechanical Problem
  • Unintended System Motion
  • No Apparent Adverse Event

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

IT WAS REPORTED THAT THE LOCK, WILL NOT REMAIN IN LOCKED POSITION DURING USE. THE BRACE HAS NOT BEEN RETURNED FOR EVALUATION. THE ROOT CAUSE OF THE COMPLAINT IS DAMAGED COMPONENT BASED ON OTHER DEVICES THAT HAVE BEEN EVALUATED. THE DEVICES LOCK BUTTON FAILED DURING MANUFACTURER TESTING CONFIRMED TO BE THE SAME MALFUNCTION FOUND IN REPORT NUMBER (B)(4). ADDITIONAL REPORTING ON THIS EVENT WILL BE PROVIDED AS A SUPPLEMENTAL REPORT TO THIS DOCUMENT IF IT BECOMES AVAILABLE.

Description of Event or Problem

IT WAS REPORTED THAT THE LOCK, WILL NOT REMAIN IN LOCKED POSITION DURING USE.