inforMED
MalfunctionKGI

HORIZON

Received Sep 6, 2024 · Event occurred Aug 21, 2024

Report 1220984-2024-00102 · MDR key 20162571

Device

Generic name

Densitometer, Bone

Manufacturer

Hologic, Inc

Model number

HORIZON-CI

Catalog number

HORIZON-CI

Product problems

  • Fire
  • Sparking
  • Fire
  • Sparking

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

H3 OTHER TEXT : OTHER.

Description of Event or Problem

IT WAS REPORTED ON AUGUST 21, THAT AFTER TURNING ON THE HORIZON DEVICE, THE TECHNICIAN HEARD A LOUD POPPING SOUND, THEN SPARKS, THEN A FLAME AND SMOKE. NO ADDITIONAL INFORMATION AVAILABLE.

Additional Manufacturer Narrative

DEVICES EVALUATION BY FIELD ENGINEER: A FIELD ENGINEER EXAMINED THE DEVICE AND COULD NOT CONFIRM IF FLAMES WERE COMING OUT OF THE MACHINE AS THERE WAS NO DAMAGE TO SURROUNDING AREAS. THE FIELD ENGINEER THE LOW-VOLTAGE POWER SUPPLY WAS REPLACED AND THE SYSTEM WAS RETURNED TO SERVICE.