inforMED
MalfunctionPWH

NRFIT CONNECTOR, ADAPTER, FILLING, MALE/FEMALE LUER, 50/BX

Received Sep 5, 2024 · Event occurred Aug 7, 2024

Report 3012307300-2024-08644 · MDR key 20152206

Device

Generic name

Administrations Sets With Neuraxial Connectors

Catalog number

21-7664-24

Lot number

6010324

Product problems

  • Fluid/Blood Leak
  • Fluid/Blood Leak

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

H3: DEVICE NOT RECEIVED BY MANUFACTURER. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE.

Description of Event or Problem

IT WAS REPORTED THAT THE ADAPTER OF THE CADD CASSETTE IS LEAKING AT THE CONNECTION ON THE MIDDLE. THERE WAS NO PATIENT INVOLVEMENT.

Additional Manufacturer Narrative

H6: EVALUATION CODES UPDATED. DEVICE EVALUATION: ONE COMPONENT WAS RETURNED FOR INVESTIGATION. THE COMPONENT WAS AIR LEAK TESTED PER ICU MEDICAL SPECIFICATIONS AND FAILED. THE COMPLAINT CAN BE CONFIRMED. A QUALITY ALERT HAS BEEN POSTED COVERING PROPER SOLVENT APPLICATION FOR ASSOCIATE REVIEW. IN THE PAST TWO YEARS, NO OTHER COMPLAINTS HAVE BEEN FILED FOR THE DESCRIBED DEFECT. ICU MEDICAL CONTINUES TO TRACK AND TREND COMPLAINTS DATA AND ESCALATE AS NEEDED. NO NON-CONFORMANCE REPORTS WERE ISSUED AGAINST THIS LOT DURING THE MANUFACTURING.