Received Sep 4, 2024 · Event occurred Aug 29, 2024
Report 9610877-2024-56701 · MDR key 20132249
Device
Generic name
Fiberopticintubationscope 1.4c 3.4tp Rop
Manufacturer
Hoya Corporation Pentax Tokyo OfficeModel number
FI-10RBSProduct problems
- Loose or Intermittent Connection
- Loose or Intermittent Connection
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE BIOPSY INLET PIECE LOOSE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE BIOPSY INLET PIECE. THIS DEFECT OCCURRED NOT DURING PROCEDURE. BASED ON THE TECHNICAL REPORT "HR-RPT-0620 (CONTROL BODY)" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
Description of Event or Problem
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. BIOPSY INLET PIECE LOOSENED.