MalfunctionLFZ

ELECSYS CMV IGM

Received Sep 3, 2024 · Event occurred Aug 9, 2024

Report 1823260-2024-02542 · MDR key 20129424

Device

Generic name

Enzyme Linked Immunoabsorbent Assay, Cytomegalovirus

Manufacturer

Roche Diagnostics

Catalog number

07027133190

Lot number

764032

Product problems

  • False Negative Result
  • False Negative Result

Patient

55 YR · Female

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

INVESTIGATION OF THE PROVIDED PATIENT SAMPLE MATERIAL FOUND THE CUSTOMER'S NON-REACTIVE ELECSYS CMV IGM RESULTS WERE REPRODUCIBLE. THE NON-REACTIVE ELECSYS CMV IGM RESULT WAS CONFIRMED BY THE RECOMLINE CMV IGM BLOT (MIKROGEN). THE INVESTIGATION DETERMINED THE SAMPLE WAS CONSISTENT WITH AN ACUTE PRIMARY CMV INFECTION (2-3 MONTHS AGO) DUE TO THE DETECTION OF LOW IGG TITERS WITH A LOW CMV IGG AVIDITY AND A DECLINING CMV IGM REACTIVITY BELOW THE BORDERLINE RANGE. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM.

Additional Manufacturer Narrative

THE COBAS E 801 MODULE SERIAL NUMBER WAS (B)(6). SAMPLE MATERIAL FROM THE PATENT WAS REQUESTED FOR INVESTIGATION.

Description of Event or Problem

THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS CMV IGM RESULTS FROM THE COBAS E 801 MODULE. THE ROCHE IGM RESULT WAS 0.3 (NEGATIVE) REPEATEDLY. THE BIOMERIEUX IGM RESULT WAS 1.72 (POSITIVE).