inforMED
MalfunctionDCF

ALBT2 TINA-QUANT ALBUMIN GEN.2

Received Aug 30, 2024 · Event occurred Aug 8, 2024

Report 1823260-2024-02515 · MDR key 20106309

Device

Generic name

Albumin Immunological Test System

Manufacturer

Roche Diagnostics

Catalog number

04469658190

Lot number

73284101

Product problems

  • Non Reproducible Results
  • Non Reproducible Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED.

Additional Manufacturer Narrative

THE ANALYZER SERIAL NUMBER IS (B)(6). THE CALIBRATION AND QC RECOVERY DATA PROVIDED SHOWED CALIBRATION FAILURES AND QC IMPRECISION. THE INVESTIGATION IS ONGOING.

Description of Event or Problem

THERE WAS AN ALLEGATION OF QUESTIONABLE ALBT2 TINA-QUANT ALBUMIN GEN.2 RESULTS FOR 1 PATIENT URINE SAMPLE ON A COBAS 6000 CORE UNIT. THE INITIAL URINE ALBUMIN RESULT WAS 182.5 MG/L. THE SAMPLE WAS REPEATED AND THE RESULT WAS 2.9 MG/L ACCOMPANIED BY A FLAG INDICATING THE VALUE IS BELOW THE LINEARITY/MEASURING RANGE OF THE ASSAY. THE SAMPLE WAS REPEATED AGAIN AND THE RESULT WAS 6.1 MG/L.