inforMED
InjuryKWI

BI POLAR RADIAL CUP 22MM DIA

Received Aug 29, 2024 · Event occurred Jul 22, 2021

Report 3000931034-2024-00512 · MDR key 20096536

Device

Generic name

Prosthesis, Elbow, Hemi-, Radial, Polymer

Manufacturer

Tornier S.a.s.

Catalog number

DTJ022

Lot number

UNKNOWN

Product problems

  • Device Dislodged or Dislocated
  • Device Dislodged or Dislocated

Patient

53 YR · Male

  • Subluxation
  • Subluxation

Narrative

Description of Event or Problem

SUBJECT: (B)(6). WELBOW CLINICAL STUDY. ONE DAY AFTER THE SURGERY INVOLVING THE CRF II IMPLANT, A CHANGE IN THE DIMENSIONS OF THE OPERATED LIMB WAS OBSERVED. IT WAS DETERMINED THAT A SHORT RADIAL BONE CUT HAD CAUSED AN ELBOW DISLOCATION. THIS EVENT WAS REPORTED AS A SERIOUS ADVERSE EVENT ¿PROLONGATION OF EXISTING HOSPITALIZATION BY 3 DAYS TO ADDRESS THIS ISSUE, REVISION SURGERY WAS REQUIRED (RE-OPERATION WITH IMPLANT CHANGE) ON (B)(6) 2024.

Additional Manufacturer Narrative

BASED ON THE AVAILABLE INFORMATION THE DEVICE WILL NOT BE RETURNED THEREFORE AN EVALUATION OF THE DEVICE CANNOT BE PERFORMED. A REVIEW OF THE DEVICE HISTORY IS NOT POSSIBLE BECAUSE THE LOT NUMBER WAS NOT COMMUNICATED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.