BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE
Received Aug 27, 2024 · Event occurred Jul 29, 2024
Report 1917413-2024-00782 · MDR key 20081356
Device
Generic name
Lymphocyte Separation Medium
Manufacturer
Becton, Dickinson & Co. (broken Bow)Catalog number
362761
Lot number
3248764
Product problems
- Incorrect, Inadequate or Imprecise Result or Readings
- Incorrect, Inadequate or Imprecise Result or Readings
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES, BUT 1 PHOTO WAS PROVIDED FOR INVESTIGATION. THE PHOTO WAS REVIEWED AND THE INDICATED FAILURE MODE FOR POOR BARRIER SEPARATION WAS OBSERVED. ADDITIONALLY, 33 RETENTION SAMPLES FROM BD INVENTORY WERE VISUALLY INSPECTED WITH NO ISSUES OBSERVED. COMPLAINTS FOR SAMPLE QUALITY ARE UNDER STATISTICAL CONTROL FOR THE MONTH OF (B)(6) 2024. AT THIS TIME, FURTHER TESTING IS NOT INDICATED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE POOR BARRIER SEPARATION BASED ON THE PHOTO PROVIDED. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. BD QUALITY WILL CONTINUE TO MONITOR SAMPLE QUALITY COMPLAINTS.
Description of Event or Problem
IT WAS REPORTED DURING USE OF BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE POOR BARRIER SEPARATION OCCURRED WITH 3 SAMPLES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.