inforMED
MalfunctionMVR

GEM MICROVASCULAR ANASTOMOTIC COUPLER

Received Aug 26, 2024 · Event occurred Jul 16, 2024

Report 1416980-2024-04502 · MDR key 20065870

Device

Generic name

Device,anastomotic,microvascular

Model number

GEM2750 I

Catalog number

511100100010

Lot number

SP23G13-1906409

Product problems

  • Detachment of Device or Device Component
  • Detachment of Device or Device Component

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

E1: INITIAL REPORTER FACILITY NAME: (B)(6). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem

IT WAS REPORTED THAT THE ANASTOMOSIS RINGS OF TWO (2) 1.0MM COUPLERS DETACHED. THE ISSUE WAS FURTHER DESCRIBED AS, "COULD NOT ACHIEVE PROPER ANASTOMOSIS". THIS OCCURRED PRIOR TO PATIENT USE. TO RESOLVE THE EVENT, THE SURGEON PERFORMED THE ANASTOMOSIS MANUALLY. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Additional Manufacturer Narrative

H11: A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. THE DEVICES WERE NOT RECEIVED FOR EVALUATION; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.