inforMED
MalfunctionLWH

SITUATE

Received Aug 20, 2024 · Event occurred Jul 19, 2024

Report 1717344-2024-01644 · MDR key 20008553

Device

Generic name

Counter, Sponge, Surgical

Model number

01-0043

Catalog number

01-0043

Product problems

  • False Positive Result
  • Device Displays Incorrect Message
  • False Positive Result
  • Device Displays Incorrect Message

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem

ACCORDING TO THE REPORTER, THE CONSOLE HAD SYSTEM ERROR. THERE WAS NO PATIENT INVOLVEMENT. MEDTRONIC'S INITIAL EVALUATION OF THE INCIDENT DEVICE FOUND THAT MULTIPLE SCANS WERE ATTEMPTED WITH THE BLAIR-PORT WAND, THE DEVICE ALERTED FOR METAL PRESENCE EACH SCAN/FALSE DETECTION .

Additional Manufacturer Narrative

EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS AVAILABLE FOR EVALUATION. FUNCTIONAL TESTING FOUND THAT MULTIPLE SCANS WERE ATTEMPTED WITH THE BLAIR-PORT WAND, THE DEVICE ALERTED FOR METAL PRESENCE EACH SCAN DUE TO A DEFECTIVE ANALOG BOARD. IT WAS REPORTED THAT THE CONSOLE HAD SYSTEM ERROR. THE REPORTED ISSUE WAS CONFIRMED. THE MOST LIKELY CAUSE WAS TRACED TO COMPONENT FAILURE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.