PATIENT CABLE
Received Aug 14, 2024 · Event occurred Jul 15, 2024
Report 2182208-2024-03357 · MDR key 19980094
Device
Generic name
Cable, Transducer And Electrode, Patient, (inc
Manufacturer
Medtronic, Inc.Model number
5492VLCatalog number
5492VL
Lot number
2208112
Product problems
- Material Integrity Problem
- Material Integrity Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
Description of Event or Problem
IT WAS REPORTED THAT THERE WAS A DIFFICULTY WITH THE PATIENT CABLE CONNECTION AND REMAINING CONNECTED. THERE WAS NO PATIENT INVOLVEMENT.