MOLECULAR SWAB COLLECTION KIT

Received Aug 12, 2024 · Event occurred Aug 5, 2024

Report MW5158365 · MDR key 19975757

Device

Generic name

Vaginitis And Bacterial Vaginosis Nucleic Acid Detection System

Model number

443925

Patient

NA · Unknown

  • Insufficient Information
  • Insufficient Information

Narrative

Description of Event or Problem

BD MOLECULAR SWAB COLLECTION KIT REF 443925-- THE SWAB WAS NOT BROKEN OFF AT THE APPROPRIATE MARK. A SWAB THAT IS TOO LONG CAN CAUSE LEAKAGE OF SAMPLE, CROSS-CONTAMINATION OR AN INSTRUMENT FAILURE. TESTING WILL STILL BE PERFORMED. THIS IS AN ONGOING ISSUE SINCE LATE 2023. REVIEW OF THESE INCIDENTS INDICATE THAT THERE IS WIDESPREAD CONFUSION WHERE TO BREAK THE SWAB. THERE IS A SOLID BLACK LINE HOWEVER THE SCORED BREAK POINT IN LOWER ON THE SWAB. RECOMMENDATION INCLUDES HAVING THE BLACK LINE REMOVED OR PLACING ON THE SCORED AREA OF THE SWAB.