NRFIT RESERVOIR CASSETTE
Received Aug 9, 2024 · Event occurred Jul 16, 2024
Report 9617604-2024-00644 · MDR key 19956191
Device
Generic name
Daministrations Sets With Neuraxial Connectors
Manufacturer
Smiths Healthcare Manufacturing S.a. De C.v.Catalog number
21-7609-24
Lot number
4448960
Product problems
- Fluid/Blood Leak
- Fluid/Blood Leak
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE.
Description of Event or Problem
IT WAS REPORTED THAT THERE WAS A LEAK WHILE FILLING THE CASSETTE. THE LEAK WAS IN THE INNER POCKET OF THE CASSETTE. THIS MAY HAVE OCCURRED DURING MANUFACTURING. THERE WAS NO PATIENT INVOLVEMENT, AND NO PATIENT HARM REPORTED.
Additional Manufacturer Narrative
ONE SAMPLE CASSETTE PRODUCT WAS RETURNED FOR EVALUATION. THE SAMPLE WAS RECEIVED IN USED CONDITIONS, DECONTAMINATED, WITHOUT ITS ORIGINAL PACKAGING AND INSIDE IN A PLASTIC BAG. NO PICTURES WERE ATTACHED TO THE COMPLAINT. FUNCTIONAL TESTING WAS PERFORMED. WHILE INJECTING LIQUID INTO THE SAMPLE, A LEAK WAS DETECTED IN THE MEDICATION BAG. ACCORDING TO SAMPLE RECEIVED, THE FAILURE MODE ¿LEAKING¿ WAS CONFIRMED IN THE DEVICE RECEIVED. LOT HISTORY REVIEW FOUND NO COMPLAINTS DOCUMENT FOR THIS LOT NUMBER.