inforMED
MalfunctionCDP

ELECSYS T3

Received Jul 30, 2024

Report 1823260-2024-02211 · MDR key 19865788

Device

Generic name

Radioimmunoassay, Total Triiodothyronine

Manufacturer

Roche Diagnostics

Model number

T3

Catalog number

ASKU

Lot number

ASKU

Product problems

  • Incorrect Measurement
  • High Test Results
  • Incorrect Measurement
  • High Test Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

1. SUMMARY OF INVESTIGATION RESULTS: THE INVESTIGATION IS ONGOING. 2. SUMMARY OF FOLLOW UP/CORRECTIVE ACTIONS TAKEN: THE INVESTIGATION IS ONGOING. 3. DEVICE RETURNED TO MANUFACTURER? NO. 4. DEVICE LABELED "FOR SINGLE USE?" NO. 5. DEVICE REPROCESSED AND REUSED? NO.

Description of Event or Problem

1. THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS T3 RESULTS FROM ONE PATIENT FROM THE COBAS E 801 ANALYTICAL UNIT. 2. PATIENT AGE RANGE: 52 YEAR. 3. PATIENT WEIGHT RANGE: ASKU. 4. PATIENT SEX BREAKDOWN: FEMALE. 5. PATIENT RACE BREAKDOWN: ASKU. 6. PATIENT ETHNICITY BREAKDOWN: ASKU.

Additional Manufacturer Narrative

1. SUMMARY OF INVESTIGATION RESULTS: THE INVESTIGATION FOUND RUTHENIUM INTERFERENCE IN THE PATIENT SAMPLE. THIS INTERFERENCE IS DOCUMENTED IN PRODUCT LABELING FOR THE ASSAY. IN RARE CASES, INTERFERENCE DUE TO EXTREMELY HIGH TITERS OF ANTIBODIES TO ANALYTE-SPECIFIC ANTIBODIES, STREPTAVIDIN OR RUTHENIUM CAN OCCUR. THESE EFFECTS ARE MINIMIZED BY SUITABLE TEST DESIGN. FOR DIAGNOSTIC PURPOSES, THE RESULTS SHOULD ALWAYS BE ASSESSED IN CONJUNCTION WITH THE PATIENT¿S MEDICAL HISTORY, CLINICAL EXAMINATION, AND OTHER FINDINGS. 2. SUMMARY OF FOLLOW UP/CORRECTIVE ACTIONS TAKEN: SAMPLE MATERIAL FROM THE PATIENT WAS SUBMITTED FOR INVESTIGATION.