inforMED
MalfunctionCGA

CARDIOCHEK PLUS ANALYZER

Received Jul 30, 2024 · Event occurred May 18, 2024

Report 1836135-2024-00027 · MDR key 19862269

Device

Generic name

Clinical Chemistry Analyzer

Manufacturer

Pts Diagnostics

Product problems

  • Electrical /Electronic Property Problem
  • Electrical /Electronic Property Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE MALFUNCTIONED, AS THE ALLEGATION OF OVERHEATING COULD NOT BE DUPLICATED. IT IS KNOWN THAT IMPROPER INSERTION OF BATTERIES - IN ANY DEVICE - MAY RESULT IN OVERHEATING, BATTERY LEAKAGE, ETC., WHICH IS STATED IN THE LABELING OF MOST BATTERY MANUFACTURERS.

Description of Event or Problem

THE CUSTOMER REPORTED THAT DURING TESTING THEIR CARDIOCHEK PLUS ANALYZER SHUT OFF AND OVERHEATED. THEY REPLACED THE BATTERIES AND THE ANALYZER OVERHEATED AGAIN. THERE WERE NO ALLEGATIONS OF PATIENT/USER HARM. THERE WERE NO ALLEGATIONS OF INCORRECT RESULTS.