K-REAMER M-ACCESS 21MM 020
Received Jul 29, 2024 · Event occurred Apr 1, 2024
Report 8031010-2024-00093 · MDR key 19850445
Device
Generic name
Reamer, Pulp Canal, Endodontic
Manufacturer
Maillefer Instruments Holding SarlCatalog number
A11MA02102012
Lot number
1745615
Product problems
- Break
- Break
Patient
NA · Unknown
- Device Embedded In Tissue or Plaque
- Device Embedded In Tissue or Plaque
Narrative
Additional Manufacturer Narrative
AS A RESULT OF THIS MALFUNCTION, THE POTENTIAL FOR SURGICAL INTERVENTION EXISTS TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION AS EVIDENCED BY PREVIOUS REPORTED EVENTS WITH SIMILAR FILES. THIS EVENT, THEREFORE, IS REPORTABLE PER 21CFR PART 803. INVESTIGATION RESULTS: SUMMARY: INVOLVED PRODUCT THAT BROKE DURING USE WAS NOT RETURNED AND CANNOT BE ANALYZED. MOREOVER, NO UNUSED FILE IS AVAILABLE FOR EVALUATION. NOTHING UNUSUAL TO REPORT WAS FOUND DURING DHR REVIEW (BATCH #1745615). ROOT CAUSES ARE NOT IDENTIFIED. WE WILL TRACK THIS KIND OF EVENT AND MONITOR THE TREND. POTENTIAL ROOT CAUSES MAY BE INCORRECT TECHNIQUE (HAND USED FILE - TECHNIQUE NO MORE VERIFIABLE TO DATE), OVERUSE (NUMBER OF USES NOT COMMUNICATED), EXCESSIVE WEAR, OR MATERIAL ISSUE (NO ANALYSIS OF THE BROKEN FRAGMENTS POSSIBLE). ALL COMPLAINTS ARE MONITORED THROUGH THE MONTHLY PRODUCT SURVEILLANCE COMMITTEE AND A CORRECTIVE ACTION COULD BE DETERMINED BY THE COMMITTEE.
Description of Event or Problem
IN THIS EVENT IT IS REPORTED THAT A K-REAMER M-ACCESS 21MM 020 FILE BROKE DURING USE. THE BROKEN PART COULD NOT BE RETRIEVED AND WAS INCORPORATED INTO THE FILL. NO INJURY.