BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE
Received Jul 25, 2024 · Event occurred Jun 26, 2024
Report 1917413-2024-00688 · MDR key 19833417
Device
Generic name
Lymphocyte Separation Medium
Manufacturer
Becton, Dickinson & Co. (broken Bow)Catalog number
362761
Lot number
2341883
Product problems
- Incorrect, Inadequate or Imprecise Result or Readings
- Incorrect, Inadequate or Imprecise Result or Readings
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
E1. INITIAL REPORTER PHONE #: (B)(6). INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES, BUT 5 PHOTOS WERE PROVIDED FOR INVESTIGATION. VISUAL EXAMINATION OF THE PHOTOS WAS PERFORMED AND REVEALED POOR BARRIER SEPARATION. THE PRODUCT EXPIRED 31DEC2023; THEREFORE, NO FURTHER TESTING CAN BE COMPLETED. COMPLAINTS FOR SAMPLE QUALITY ARE UNDER STATISTICAL CONTROL FOR THE MONTH OF JUNE 2024. AT THIS TIME, FURTHER TESTING IS NOT INDICATED AND TUBES EXPIRED DECEMBER 2023. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR POOR BARRIER SEPARATION BASED ON THE PHOTOS PROVIDED. AN EXACT ROOT CAUSE COULD NOT BE DETERMINED. THE INSTRUCTION FOR USE (IFU) CAUTIONS: DO NOT USE TUBES AFTER EXPIRATION DATE PRINTED ON THE TUBE LABEL. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. BD QUALITY WILL CONTINUE TO MONITOR SAMPLE QUALITY COMPLAINTS.
Description of Event or Problem
IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE AN UNSPECIFIED NUMBER OF TUBES FAILED TO SEPARATE. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Additional Manufacturer Narrative
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D9: DEVICE AVAILABLE FOR EVALUATION: YES H.3 DEVICE EVAL BY MANUFACTURER? YES D9: RETURNED TO MANUFACTURER ON: 04-SEP-2024. INVESTIGATION SUMMARY BD RECEIVED EXPIRED SAMPLES ON (B)(6) 2024. 5 PHOTOS WERE ALSO PROVIDED FOR INVESTIGATION. VISUAL EXAMINATION OF THE PHOTOS WAS PERFORMED AND REVEALED POOR BARRIER SEPARATION. THE EXPIRED SAMPLES WERE VISUALLY INSPECTED WITH NO ISSUES OBSERVED. THE PRODUCT EXPIRED 31DEC2023; THEREFORE, NO FURTHER TESTING CAN BE COMPLETED. COMPLAINTS FOR SAMPLE QUALITY ARE UNDER STATISTICAL CONTROL FOR THE MONTH OF (B)(6) 2024. AT THIS TIME, FURTHER TESTING IS NOT INDICATED AND TUBES EXPIRED DECEMBER 2023. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR POOR BARRIER SEPARATION BASED ON THE PHOTOS PROVIDED. AN EXACT ROOT CAUSE COULD NOT BE DETERMINED. THE INSTRUCTION FOR USE (IFU) CAUTIONS: DO NOT USE TUBES AFTER EXPIRATION DATE PRINTED ON THE TUBE LABEL. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. BD QUALITY WILL CONTINUE TO MONITOR SAMPLE QUALITY COMPLAINTS.
Description of Event or Problem
IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE AN UNSPECIFIED NUMBER OF TUBES FAILED TO SEPARATE. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.