inforMED
MalfunctionGMS

ANTI-HUMAN GLOBULIN (RABBIT/MURINE MONOCLONAL)

Received Jan 17, 2011 · Event occurred Oct 15, 2010

Report 9610824-2011-00005 · MDR key 1983143

Device

Generic name

Anti-human Globulin, Anti-igg, -c3d, Poly.

Catalog number

804115100

Lot number

7836090-02

Product problems

  • False Positive Result

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

THE CUSTOMER REPORTED THAT CAP SURVEY (B)(4) FAILED TO REACT (B)(6) WITH ANTI-HUMAN GLOBULIN ANTI-IGG, -C3D, POLYSPECIFIC (B)(4), LOT 7836090-02. THE CUSTOMER HAS SENT US THE SURVEY SAMPLE AND THE COMPLAINED REAGENT. BOTH WERE TESTED IN QUALITY CONTROL LABORATORY. THE SURVEY SAMPLE REACTED CORRECTLY (B)(6) WITH THE AFFECTED LOT OF ANTI-HUMAN GLOBULIN ANTI-IGG, -C3D, POLYSPECIFIC.