inforMED
MalfunctionPTS

GRASPING FORCEPS

Received Jul 25, 2024 · Event occurred Jul 16, 2024

Report 9614641-2024-01507 · MDR key 19831145

Device

Generic name

Grasping Forceps

Model number

FG-53SX-1

Lot number

K3525-6506

Product problems

  • Break
  • Break

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE DEVICE EVALUATION IS ONGOING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem

IT WAS OBSERVED DURING THE DEVICE INSPECTION, THE GRASPING FORCEPS PARTS BROKE OFF. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

Additional Manufacturer Narrative

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE LEGAL MANUFACTURER'S FINAL INVESTIGATION, ASSOCIATED WITH UPDATE TO H3, H6 CODING. BASED ON THE INVESTIGATION RESULT ,IT IS LIKELY THAT A FORCE WAS APPLIED TO THE LINKAGE MECHANISM WHILE HANDLING THE DEVICE. AS A RESULT, THE LINKAGE MECHANISM PART WHERE DECREASED ITS STRENGTH WAS BROKEN AND FELL OFF. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. THE EVENT CAN BE DETECTED/PREVENTED BY FOLLOWING THE INSTRUCTIONS FOR USE (IFU) WHICH STATE THE FOLLOWING: BEFORE USE, CONFIRM THAT THE INSTRUMENT IS NOT CORRODED, DENTED OR DISCOLORED ; DO NOT USE IT IF ANY OF THESE CONDITIONS ARE OBSERVED. IF THE INSTRUMENT IS DAMAGED, PART OF IT COULD FALL OFF INSIDE THE PATIENT, OR IT COULD BECOME IMPOSSIBLE TO OPEN OR CLOSE. IF THE INSTRUMENT DOES NOT OPERATE PROPERLY DURING THE PROCEDURE, OR IF IT BECOMES DIFFICULT TO OPEN OR CLOSE THE GRASPING JAWS, STOP THE PROCEDURE IMMEDIATELY AND WITHDRAW THE INSTRUMENT INTO THE ENDOSCOPE¿S CHANNEL. IF THIS IS NOT POSSIBLE, CAREFULLY REMOVE IT TOGETHER WITH THE ENDOSCOPE TO AVOID CAUSING PATIENT INJURY. REPROCESS THE INSTRUMENT IMMEDIATELY AFTER USE, FIRST BY IMMERSING IT IN A NEUTRAL, LOW-FOAMING, MEDICAL GRADE DETERGENT SOLUTION, THEN FOLLOWING THE CLEANING AND STERILIZATION PROCEDURES GIVEN IN THIS CHAPTER. FAILURE TO REPROCESS THE INSTRUMENT IMMEDIATELY AFTER USE, OR USING OTHER THAN A MEDICAL-GRADE DETERGENT MAY CAUSE CORROSION AT THE METAL PARTS LIKE THE GRASPING JAWS. THIS COULD IMPAIR THE OPERATION OF THE INSTRUMENT OR CAUSE A PART OF IT TO BREAK AND/OR FALL OFF INSIDE THE PATIENT. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.