UMBILICAL CATHETER DUAL LUMEN 5FR
Received Jul 24, 2024 · Event occurred Jul 16, 2024
Report 1423537-2024-00086 · MDR key 19821948
Device
Generic name
Catheter, Umbilical Artery
Manufacturer
Cardinal HealthModel number
8888160556Catalog number
8888160556
Product problems
- Connection Problem
- Connection Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE DEVICE HISTORY RECORD REVIEW COULD NOT BE PERFORMED SINCE NO LOT NUMBER WAS PROVIDED. THE DEVICE WAS RECEIVED FOR EVALUATION AND THE REPORTED CONDITION WAS OBSERVED. A CORRECTIVE AND PREVENTIVE ACTION HAS BEEN INITIATED TO FURTHER INVESTIGATE THIS ISSUE.
Additional Manufacturer Narrative
AN INVESTIGATION IS CURRENTLY UNDERWAY. UPON COMPLETION, THE RESULTS WILL BE FORWARDED.
Description of Event or Problem
THE CUSTOMER REPORTED THAT WHAT USED TO BE ROUNDED EDGES AT THE END OF THE PORTS NOW HAVE (WHAT THEY WILL CALL) AS TABS ON THE RIGHT AND LEFT SIDES. IF ONE OR BOTH OF THESE TABS GETS WORN DOWN, THE STOPCOCK THAT THEY ROUTINELY CHANGE DOES NOT SCREW ON SECURELY ANYMORE, RESULTING IN PATIENT SAFETY RISK AND THE UMBILICAL LINE HAD TO BE REMOVED AND A NEW TYPE OF INTRAVENOUS ACCESS OBTAINED. THIS UMBILICAL LINE WAS ONLY IN USE FOR 5 DAYS (SHOULD LAST UP TO 14 DAYS OR LONGER) BEFORE THIS ISSUE AROSE. THERE WAS NO PATIENT HARM REPORTED.