inforMED
DeathBTM

AMBU SPUR II ADULT TUBE RES., M MASK

Received Jul 23, 2024 · Event occurred May 25, 2024

Report 1220828-2024-00014 · MDR key 19810671

Device

Generic name

Ambu Spur Ii Adult Tube Res., M Mask

Manufacturer

Ambu Inc.

Model number

521000001

Lot number

1000919409

Product problems

  • Crack
  • Crack

Patient

NA · Unknown

  • Cardiac Arrest
  • Cardiac Arrest

Narrative

Additional Manufacturer Narrative

THE REPORTED FAILURE OF COMPRESSIBLE BAG NOT RE-INFLATING AND HAVING A CRACK OR SEPARATION, COULD NOT BE VERIFIED AS NO SAMPLE OR PHOTO WAS RETURNED FOR INVESTIGATION. REVIEW OF PRODUCTION RECORDS DID NOT REVEAL ANY ABNORMALITY. POSSIBLE CAUSES FOR COMPRESSIBLE BAG NOT RE-INFLATING ARE A DEFORMATION OF THE COMPRESSIBLE BAG OR A MALFUNCTION OF INLET VALVE DISC. POSSIBLE CAUSES FOR A CRACK OR SEPARATION ARE A DAMAGE BY A SHARP OBJECT DURING USE OR A DETACHMENT OF INLET VALVE OR PATIENT VALVE HOUSING FROM THE COMPRESSIBLE BAG. THERE ARE 100% FUNCTION TESTS PERFORMED AFTER MANUFACTURING, INCLUDING A VERIFICATION OF RE-EXPANSION TO THE DEFAULT STATE AFTER COMPRESSION OF SPUR II. COMPRESSIBLE BAG CRACK OR INABILITY TO REINFLATE WOULD BE DETECTED DURING VISUAL INSPECTION AS A PART OF FINAL QUALITY CONTROL PROCEDURE. SPUR II IS PACKED MANUALLY ACCORDING TO PACKING PROCEDURE. THUS, THE PRODUCT SHOULD BE WITHIN SPECIFICATIONS PRIOR TO DELIVERY TO CUSTOMER. DUE TO THE LIMITED INFORMATION, THE ROOT CAUSE COULD NOT BE ESTABLISHED. AS STATED IN THE INSTRUCTIONS FOR USE (LBL-005931 REV.15) "ALWAYS INSPECT THE RESUSCITATOR AND ACCESSORIES (E.G. FACE MASK, PEEP VALVE, FILTERS, ETC.) AND PERFORM A FUNCTIONAL TEST AFTER UNPACKING, CLEANING, ASSEMBLY AND PRIOR TO USE. WE WILL KEEP MONITORING SIMILAR INCIDENTS.

Description of Event or Problem

DURING THE COURSE OF CARDIAC ARREST RESUSCITATION, THE BAG PORTION OF THE BAG VALVE MASK ASSEMBLY FAILED TO RE-INFLATE. UPON ARRIVAL AT THE HOSPITAL, A MEMBER OF THE EMERGENCY DEPARTMENT STAFF LOCATED A SEPARATION IN OR CRACK IN THE RUBBER BAG. PATIENT WAS PRONOUNCED DECEASED AFTER ARRIVING AT THE HOSPITAL.