ELECSYS PROBNP II
Received Jul 16, 2024 · Event occurred Jun 25, 2024
Report 1823260-2024-02043 · MDR key 19749818
Device
Generic name
Natriuretic Peptide Test
Manufacturer
Roche DiagnosticsCatalog number
09315284214
Lot number
ASKU
Product problems
- Low Test Results
- Low Test Results
Patient
NA · Male
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS PROBNP II RESULTS COMPARED TO BNP FOR TWO SAMPLES FROM ONE PATIENT USING A COBAS E 801 ANALYTICAL UNIT. THE CLINICIAN DETERMINED THE RESULTS WERE NOT CONSISTENT WITH THE CLINICAL FINDINGS. SAMPLE 1 WAS COLLECTED ON (B)(6) 2024. THE PROBNP RESULT WAS <5 PG/ML AND THE BNP RESULT BY ANOTHER UNKNOWN METHOD WAS 697 PG/ML. ON (B)(6) 2024, THE INITIAL INVESTIGATION OF THE SAMPLE USING ANOTHER COBAS E801 AND REAGENT LOT NUMBER 732327 HAD A PROBNP RESULT OF <5 PG/ML. SAMPLE 2 WAS COLLECTED ON (B)(6) 2024. THE PROBNP RESULT WAS <5 PG/ML AND THE BNP RESULT BY ANOTHER UNKNOWN METHOD WAS 456 PG/ML. ON (B)(6) 2024, THE INITIAL INVESTIGATION OF THE SAMPLE USING ANOTHER COBAS E801 AND REAGENT LOT NUMBER 732327 HAD A PROBNP RESULT OF <5 PG/ML.
Additional Manufacturer Narrative
THE COBAS E 801 ANALYTICAL UNIT SERIAL NUMBER WAS (B)(6). SAMPLE FROM THE PATIENT WAS REQUESTED FOR FURTHER INVESTIGATION.
Additional Manufacturer Narrative
SAMPLE FROM THE PATIENT WAS INVESTIGATED AND THE CUSTOMER¿S RESULT WAS CONFIRMED. FURTHER TESTING REVEALED THE SAMPLE CONTAINED TWO POINT MUTATIONS (R72H AND E69D) IN THE EPITOPE OF THE DETECTION ANTIBODY AND WAS THEREFORE NOT DETECTED BY ELECSYS PROBNP II ASSAY. PER PRODUCT LABELING FOR THE ASSAY: IN EXTREMELY RARE CASES (GLOBAL INCIDENCE: < 1 IN 10 MILLION), PATIENTS MAY SHOW DISCREPANT RESULTS WHEN TESTED WITH THE ASSAY KIT (VALUES < LIMIT OF DETECTION) DUE TO A NT-PROBNP GENETIC VARIANT. FOR DIAGNOSTIC PURPOSES, THE RESULTS SHOULD ALWAYS BE ASSESSED IN CONJUNCTION WITH THE PATIENT¿S MEDICAL HISTORY, CLINICAL EXAMINATION, AND OTHER FINDINGS.