CAPTURA MINI BIOPSY FORCEPS W/O SPIKE

Received Jul 16, 2024 · Event occurred Jun 12, 2024

Report 1037905-2024-00387 · MDR key 19749811

Device

Generic name

Pts Endoscopic Grasping/cutting Instrument, Non-powered, Exempt

Model number

G53006

Catalog number

DBF-1.8-160-S

Lot number

W4831535

Product problems

  • Nonstandard Device
  • Difficult to Open or Close
  • Nonstandard Device
  • Difficult to Open or Close

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE INVESTIGATION IS ON-GOING. A FOLLOW-UP EMDR WILL BE PROVIDED WITHIN 30 DAYS OF SUBMISSION OF THIS REPORT.

Description of Event or Problem

DURING AN UNKNOWN ENDOSCOPIC PROCEDURE, THE PHYSICIAN USED TWO (2) COOK CAPTURA MINI BIOPSY FORCEPS W/O SPIKE. IT WAS REPORTED THAT THESE TWO FORCEPS BROKE DURING PROCEDURES TODAY. THERE WAS NO REPORTABLE INFORMATION AT THAT TIME. THE DEVICE WAS RETURNED ON 21JUN2024 FOR EVALUATION. THE QUALITY ENGINEERING INITIAL ASSESSMENT STATED THAT THE DEVICE RETURNED WITH THE CUPS OPENED AND WOULD NOT CLOSE WHEN HANDLE WAS MANIPULATED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Additional Manufacturer Narrative

INVESTIGATION EVALUATION: THE PRODUCTS SAID TO BE INVOLVED WERE RETURNED IN TWO OPEN POUCHES FROM THE LOT NUMBER PROVIDED IN THE REPORT. THE LABELS MATCH THE PRODUCTS RETURNED. OUR LABORATORY EVALUATION OF THE PRODUCT SAID TO BE INVOLVED CONFIRMED THE REPORT. BOTH DEVICES WERE RETURNED WITH NO SIGN OF DAMAGE TO THE HANDLES OR CATHETERS, HOWEVER THE CUPS WERE IN THE OPEN POSITION ON BOTH DEVICES. WHEN THE HANDLE WAS MANIPULATED, THE CUPS WOULD NOT CLOSE AND WERE STUCK OPEN. UNDER MAGNIFICATION IT WAS NOTED THAT THE CUPS WERE DAMAGED AND WOULD NOT CLOSE FULLY. THE DEVICES WERE RETURNED TO THE SUPPLIER FOR FURTHER EVALUATION AND THE FOLLOWING WAS PROVIDED, "BOTH RETURNED DEVICES WERE VISUALLY EVALUATED AND NO VISIBLE SIGNS OF DAMAGE WERE NOTED IN THE JAW ASSEMBLY, COIL CABLE, PEBAX COATING, OR HANDLE COMPONENTS OF EITHER DEVICE. THE DEVICES WERE EVALUATED FOR FUNCTIONALITY IN THE U-BEND CONFIGURATION, AND FAILED TO ACTUATE WHEN MANIPULATED WITH THE HANDLE. THE SPOOLS ON EACH DEVICE FULLY BOTTOMED OUT AGAINST THE THUMB LOOP, INDICATING A BREAKAGE IN THE INNER JAW/CONTROL WIRE ASSEMBLY/ THE JAWS COULD BE MANUALLY OPENED AND CLOSED USING THUMB AND INDEX FINGER AT THE DISTAL TIP, BUT JAWS COULD NOT BE ACTUATED WITH THE HANDLE. DUE TO THE NON-FUNCTIONING STATE OF BOTH DEVICES, THE SOLDER JOINT WAS FIRST TO BE EVALUATED. EACH DEVICE WAS CUT SEVERAL INCHES FROM THE DISTAL TIP USING SIDE CUTTERS. THE COIL CABLE WAS THEN CAREFULLY CUT USING A DREMEL TOOL FITTED WITH A 409 CUTOFF WHEEL. AFTER REMOVING A SEGMENT OF CABLE FROM EACH DEVICE, IT WAS DISCOVERED THAT THE BREAKAGE OCCURRED AT THE IJA-TO-CONTROL WIRE SOLDER JOINT IN BOTH DEVICES. THE CAUSE OF THIS BREAKAGE WAS DUE TO EXCESSIVE LINK WIRE MATERIAL BEING REMOVED DURING THE SOLDER JOINT GREENWHEELING PROCESS. IT IS NOTED IN THE PROCEDURE THAT THE OPERATOR SHOULD BE CAREFUL NOT TO GREENWHEEL THE LINK WIRES; THIS WAS DETERMINED TO BE THE ROOT CAUSE OF THE NONCONFORMANCES." THE DEVICE HISTORY RECORDS WERE REVIEWED; THEY WERE MANUFACTURED MARCH 2024. THERE WERE NO RELEVANT DEFECTS NOTED IN THE MANUFACTURING/FQC CHECKLISTS. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: OUR LABORATORY EVALUATION OF THE RETURNED DEVICES CONFIRMED THE REPORT. THE SUPPLIER PROVIDED THE FOLLOWING, "THE COMPLAINT WAS CONFIRMED. A REVIEW OF THE DEVICE HISTORY RECORD DID NOT REVEAL ANY ANOMALIES. THERE WERE NO SIGNS OF DAMAGE NOTED DURING THE VISUAL EVALUATION OF THE DEVICES. DURING THE FUNCTIONAL EVALUATION IT WAS NOTED THAT THE DEVICES COULD NOT BE ACTUATED WITH THE HANDLE. FURTHER EVALUATION FOUND THAT EXCESSIVE LINK MATERIAL WAS REMOVED DURING THE SOLDER JOINT GREENWHEELING PROCESS. THE CUSTOMERS COMPLAINT OF THE DEVICES BEING 'BROKE' WAS CONFIRMED. ROOT CAUSE WAS DETERMINED TO BE HUMAN ERROR. AWARENESS TRAINING WILL BE PERFORMED WITH THE OPERATORS." PRIOR TO DISTRIBUTION, ALL CAPTURA BIOPSY FORCEPS ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TEST TO ENSURE PROPER WORKABILITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A REVIEW OF THE COMPLAINT HISTORY WAS CONDUCTED. THE LIKELIHOOD OF OCCURRENCE IS CONSIDERED RARE. CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.