3-POINT SHOULDER DISTRACTION SYSTEM

Received Jul 15, 2024 · Event occurred Jun 19, 2024

Report 1220246-2024-06740 · MDR key 19740324

Device

Generic name

Traction Apparatus

Manufacturer

Arthrex, Inc.

Catalog number

AR-1600M

Lot number

14922871

Product problems

  • Detachment of Device or Device Component
  • Detachment of Device or Device Component

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

ON 06/19/2024, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA SEMS- (B)(4) THAT (2) AR-1600M 3-POINT SHOULDER DISTRACTION SYSTEMS CABLES PLASTIC WAS COMING OFF. THIS WAS DISCOVERED DURING A CASE WITH NO EFFECT ON THE PATIENT.

Additional Manufacturer Narrative

INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED UPON AVAILABILITY OF ADDITIONAL INFORMATION.

Additional Manufacturer Narrative

ADDITIONAL INFORMATION: D9, G3, H3, H6. THE COMPLAINT ALLEGATION WAS CONFIRMED. ONE UNPACKAGED AR-1600M SERIAL/BATCH NUMBER (B)(6) WAS RECEIVED FOR EVALUATION. FUNCTIONAL TESTING WAS NOT PERFORMED AS THE DEVICE WAS DAMAGED. VISUAL INSPECTION NOTED SCRATCHES ALL OVER THE DEVICE, THE CABLE COVER WAS RIPPED OFF, AND SIGNS OF WEAR AND TEAR ALL OVER THE DEVICE WERE NOTED. THE MOST LIKELY REASON FOR THE REPORTED FAILURE IS THE WEAR AND TEAR DAMAGE INCURRED OVER REPEATED USAGE.