inforMED
MalfunctionFSR

DUO HEADLIGHT 2 BAY SYSTEM - US

Received Jul 7, 2024

Report 3006697299-2024-00098 · MDR key 19686889

Device

Generic name

Pfm05

Catalog number

90620US

Product problems

  • Insufficient Heating
  • Overheating of Device
  • Insufficient Heating
  • Overheating of Device

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE DUO HEADLIGHT 2 BAY SYSTEM (90620US) FLICKERS DUE TO TOP COOLING FAN NOT WORKING. THE UNIT WOULD OPERATE NORMALLY FOR THE FIRST TEN MINUTES THEN PROCEEDED TO FLICKER WHEN THE MODULE GOT HOT. NO PATIENT INJURY, DEATH OR SURGICAL DELAY HAS BEEN REPORTED.

Additional Manufacturer Narrative

THE DUO HEADLIGHT 2 BAY SYSTEM (90620US) WAS RETURNED FOR EVALUATION: FAILURE ANALYSIS - THE DUO HEADLIGHT WAS RECEIVED IN USED CONDITION. EVALUATION IDENTIFIED THAT THE HEADLIGHT POWER CORD CONNECTOR PIN WAS BROKEN OFF. THE POWER CORD HAS BEEN REPLACED TO CORRECT THE ISSUE. THE UNIT PASSED ALL TESTS. ROOT CAUSE ANALYSIS - IT WAS DETERMINED THAT THE DAMAGE WAS MOST LIKELY DUE TO ROUGH HANDLING/ENVIRONMENTAL DAMAGE. NO MANUFACTURING, WORKMANSHIP, OR MATERIAL DEFICIENCY HAS BEEN IDENTIFIED.

Additional Manufacturer Narrative

UPDATED FIELDS G3, G6, H2, H11. CORRECTED FIELD: H6 (MEDICAL DEVICE PROBLEM CODE).

Description of Event or Problem

N/A.