inforMED
MalfunctionFLE

AMSCO 400 STERILIZER

Received Jul 3, 2024 · Event occurred Jun 4, 2024

Report 3005899764-2024-00030 · MDR key 19672099

Device

Generic name

Sterilizer

Product problems

  • Temperature Problem
  • Temperature Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A STERIS SERVICE TECHNICIAN ARRIVED ONSITE TO INSPECT THE STERILIZER AND FOUND THE HEATING ELEMENT WAS DAMAGED AND REQUIRED REPLACEMENT. THE TECHNICIAN REPLACED THE HEATING ELEMENT, TESTED THE UNIT, CONFIRMED IT TO BE OPERATING ACCORDING TO SPECIFICATION, AND RETURNED IT TO SERVICE. A 3-YEAR COMPLAINT REVIEW INDICATES THIS TO BE AN ISOLATED EVENT. A FOLLOW UP MDR SHALL BE SUBMITTED SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.

Description of Event or Problem

THE USER FACILITY REPORTED THEIR AMSCO 400 STERILIZER MADE A "POPPING NOISE" AND HAD "SMOKE" COMING FROM THE BOTTOM OF THE UNIT. THE UNIT WAS REMOVED FROM SERVICE TO AWAIT ONSITE INSPECTION. NO REPORT OF INJURY.