inforMED
MalfunctionPID

ESOFLIP

Received Jun 27, 2024 · Event occurred Feb 26, 2024

Report 9681384-2024-00093 · MDR key 19625500

Device

Generic name

Esophageal Dilator Balloon With Or Without Ele

Model number

ES-330

Catalog number

ES-330

Lot number

23A0109JZ

Product problems

  • Device Difficult to Program or Calibrate
  • Device Difficult to Program or Calibrate

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. ONE REPRESENTATIVE SAMPLE WAS AVAILABLE FOR EVALUATION. FU NCTIONAL TESTING OF THE RETURN DEVICE CONFIRMED A MANUFACTURING FAULT. THE ROOT CAUSE OF THE OBSERVED CONDITION WAS DETERMINED TO BE A RESULT OF A MANUFACTURING ACTIVITY. A PROCESS IMPROVEMENT HAS BEEN INITIATED TO PREVENT THIS CONDITION FROM RECURRING. A REVIEW OF THE DEVICE HISTORY RECORD INDICATES THE PRODUCT WAS RELEASED MEETING ALL MEDTRONIC QUALITY RELEASE SPECIFICATIONS AT THE TIME OF MANUFACTURE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem

ACCORDING TO THE REPORTER, THE CATHETER FAILED PRECHECK. EACH SENSOR WAS STUCK ON 7.6 DURING PRECHECK AND DID NOT MOVE EVEN THOUGH THE FILL VOLUME WAS CHANGING. MEDTRONIC¿S INITIAL EVALUATION OF THE INCIDENT DEVICE FOUND THAT IF THE MIXING OF SALINE SOLUTION DURING THE MANUFACTURING PROCESS IS NOT HOMOGENEOUS, THE SALINE SOLUTION WILL GIVE A CONDUCTIVITY OUT OF RANGE. THERE WAS NO PATIENT INVOLVED.