HBA1C TQ GEN.3
Received Jun 26, 2024 · Event occurred May 28, 2024
Report 1823260-2024-01868 · MDR key 19617776
Device
Generic name
Glycosylated Hemoglobin Assay
Manufacturer
Roche DiagnosticsCatalog number
05336180190
Lot number
66305801
Product problems
- False Positive Result
- Incorrect, Inadequate or Imprecise Result or Readings
- High Test Results
- Non Reproducible Results
- False Positive Result
- Incorrect, Inadequate or Imprecise Result or Readings
- High Test Results
- Non Reproducible Results
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
DATA WAS REQUESTED FOR INVESTIGATION BUT NOT PROVIDED DESPITE SEVERAL REMINDERS WITHIN MORE THAN 3 MONTHS. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.
Additional Manufacturer Narrative
THE COBAS C111 ANALYZER SERIAL NUMBER WAS (B)(6). CALIBRATION AND QC WERE ACCEPTABLE. THE INVESTIGATION IS ONGOING.
Description of Event or Problem
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 7 PATIENTS' SAMPLES TESTED WITH HBA1C ASSAY ON A COBAS C111 ANALYZER. SAMPLE 1 AND SAMPLE 2 WERE REPEATED ON (B)(6) 2024: SAMPLE 1 (PATIENT 1): INITIAL RESULT: 7.6 %. REPEAT RESULT: 5.8 %. SAMPLE 2 (PATIENT 2): INITIAL RESULT: 8.9 %. REPEAT RESULT: 6.5 %. SAMPLE 3 (PATIENT 3) WAS REPEATED ON (B)(6) 2024: INITIAL RESULT: 7.0 %. REPEAT RESULT: 5.4 %. SAMPLE 4, SAMPLE 5, SAMPLE 6, AND SAMPLE 7 WERE REPEATED ON (B)(6) 2024: SAMPLE 4 (PATIENT 4): INITIAL RESULT: 6.7 %. REPEAT RESULT: 5.2 %. SAMPLE 5 (PATIENT 5): INITIAL RESULT: 8.8 %. REPEAT RESULT: 6.7 %. SAMPLE 6 (PATIENT 6): INITIAL RESULT: 9.4 %. REPEAT RESULT: 7.4 %. SAMPLE 7 (PATIENT 7): INITIAL RESULT: 6.6 %. REPEAT RESULT: 5.2 %. THE PATIENTS QUESTIONED THEIR INITIAL RESULTS AND, THEREFORE, THE SAMPLES WERE REPEATED. THE REPEAT RESULTS WERE DEEMED TO BE CORRECT.