CADD MEDICATION CASSETTE RESERVOIR
Received Jun 21, 2024 · Event occurred May 27, 2024
Report 3012307300-2024-05318 · MDR key 19589059
Device
Generic name
Administrations Sets With Neuraxial Connectors
Manufacturer
Smiths Medical Asd, IncCatalog number
21-7609-24
Lot number
4420476
Product problems
- Fluid/Blood Leak
- Fluid/Blood Leak
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
ONE UNIT WAS RECEIVED, AND THREE PICTURES WERE ATTACHED TO THE COMPLAINT. NO DISCREPANCIES WERE DETECTED IN THE PICTURES. THE SAMPLE CONSISTS OF ONE CASSETTE PRODUCT FROM PART NUMBER 21-7609-24, LOT NUMBER 4420476. THE SAMPLE WAS RECEIVED IN USED CONDITIONS, WITHOUT ITS ORIGINAL PACKAGING AND INSIDE IN A PLASTIC BAG. THE SAMPLE WAS VISUALLY INSPECTED, AND NO DAMAGE WAS OBSERVED. FUNCTIONAL TESTING DETECTED A LEAK IN THE LUER AND TUBE. THE INVESTIGATION CONFIRMED THE CUSTOMER REPORTED ISSUE. THERE IS NOT NON-CONFORMANCE OR DEVIATION RELATED TO THE FAILURE REPORTED IN THE DEVICE HISTORY RECORD.
Additional Manufacturer Narrative
INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
Description of Event or Problem
IT WAS REPORTED THAT THERE WAS A MEDICATION LEAK DURING THE MEDICATION FILLING PROCESS. THERE WAS NO PATIENT INVOLVEMENT.