REPL PRT AR-1600M HUB LEVER
Received Jun 19, 2024 · Event occurred Apr 15, 2021
Report 1220246-2024-06154 · MDR key 19568151
Device
Generic name
Traction Apparatus
Manufacturer
Arthrex, Inc.Model number
REPL PRT AR-1600M HUB LEVERCatalog number
AR-16001408
Lot number
UNK
Product problems
- Break
- Break
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THIS 3500A RECORD IS SUBMITTED TO COMPLY WITH AN FDA 483 INSPECTIONAL OBSERVATION ISSUED TO ARTHREX INC ON MAY 5, 2023. ARTHREX HAS REASSESSED THE REPORTABILITY DECISIONS MADE ON HISTORICAL COMPLAINT RECORDS USING REVISED CRITERION. THIS 3500A DOCUMENT IS A RESULT OF THE REASSESSMENT. COMPLAINT CONFIRMED. VISUAL EVALUATION CONFIRMS THE DISTAL END OF THE HANDLE IS BROKEN AND MISSING A PIECE. THE BROKEN FRAGMENT WAS NOT RETURNED. NO FURTHER ABNORMALITY OBSERVED. THE MOST LIKELY CAUSE FOR THE REPORTED FAILURE CAN BE ATTRIBUTED TO MISUSE/MISHANDLING DUE TO DAMAGE TO THE DEVICE.
Description of Event or Problem
IT WAS REPORTED THAT A PIECE OF THE AR-16001408 HANDLE HAS BROKEN OFF AND THERE IS A SHARP PIECE OF HANDLE STICKING OUT SO IT IS NOT USABLE.