MINI JOINT DISTRACTOR
Received Jun 19, 2024 · Event occurred Apr 7, 2021
Report 1220246-2024-06152 · MDR key 19568138
Device
Generic name
Traction Apparatus
Manufacturer
Arthrex, Inc.Model number
MINI JOINT DISTRACTORCatalog number
AR-8950JD
Lot number
051845
Product problems
- Break
- Break
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THIS 3500A RECORD IS SUBMITTED TO COMPLY WITH AN FDA 483 INSPECTIONAL OBSERVATION ISSUED TO ARTHREX INC ON MAY 5, 2023. ARTHREX HAS REASSESSED THE REPORTABILITY DECISIONS MADE ON HISTORICAL COMPLAINT RECORDS USING REVISED CRITERION. THIS 3500A DOCUMENT IS A RESULT OF THE REASSESSMENT. COMPLAINT CONFIRMED, THE THREADED SHAFT BROKE. THE THREADED SHAFT WAS MEASURED AND FOUND TO MEET DIMENSIONAL SPECIFICATIONS. THE MOST LIKELY CAUSE FOR THE REPORTED FAILURE CAN BE ATTRIBUTED TO WEAR AND TEAR.
Description of Event or Problem
IT WAS REPORTED THAT DURING A LAPIDUS SURGERY THE THREADED PART BROKE WHEN THE SURGEON HAD OPENED THE DEVICE. IT WAS FURTHER REPORTED THAT NOTHING BROKE OFF AND THE PROCEDURE WAS FINISHED SUCCESSFULLY WITH ANOTHER DEVICE. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.