GRASPING FORCEPS
Received Jun 14, 2024 · Event occurred May 30, 2024
Report 9614641-2024-01260 · MDR key 19543494
Device
Generic name
Grasping Forceps
Manufacturer
Aomori Olympus Co., Ltd.Model number
FG-53SX-1Lot number
K2517
Product problems
- Break
- Break
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. IT HAS BEEN OVER 2 YEARS SINCE THE SUBJECT DEVICE WAS MANUFACTURED. BASED ON THE RESULTS OF THE INVESTIGATION, A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED. HOWEVER, IT IS LIKELY THAT "FORCEPS WAS BROKEN AND FORCEPS JAW/COMPONENT FELL OFF " OCCURRED DUE TO: FOREIGN MATERIALS OR CLEANING SOLUTION HAD LEFT ON THE ARM OF THE GRASPING JAWS DUE TO INSUFFICIENT CLEANING OF THE DEVICE. RESIDUES ON THE ARM OF THE GRASPING JAWS CORRODED AND DECREASED ITS STRENGTH. A FORCE WAS APPLIED TO THE LINKAGE MECHANISM WHILE HANDLING THE DEVICE. AS A RESULT, THE LINKAGE MECHANISM PART WHERE DECREASED ITS STRENGTH WAS BROKEN AND FELL OFF. THE EVENT CAN BE DETECTED/PREVENTED BY FOLLOWING THE INSTRUCTIONS FOR USE WHICH STATE: BEFORE USE, CONFIRM THAT THE INSTRUMENT IS NOT CORRODED, DENTED OR DISCOLORED ; DO NOT USE IT IF ANY OF THESE CONDITIONS ARE OBSERVED. IF THE INSTRUMENT IS DAMAGED, PART OF IT COULD FALL OFF INSIDE THE PATIENT, OR IT COULD BECOME IMPOSSIBLE TO OPEN OR CLOSE. IF THE INSTRUMENT DOES NOT OPERATE PROPERLY DURING THE PROCEDURE, OR IF IT BECOMES DIFFICULT TO OPEN OR CLOSE THE GRASPING JAWS, STOP THE PROCEDURE IMMEDIATELY AND WITHDRAW THE INSTRUMENT INTO THE ENDOSCOPE¿S CHANNEL. IF THIS IS NOT POSSIBLE, CAREFULLY REMOVE IT TOGETHER WITH THE ENDOSCOPE TO AVOID CAUSING PATIENT INJURY. REPROCESS THE INSTRUMENT IMMEDIATELY AFTER USE, FIRST BY IMMERSING IT IN A NEUTRAL, LOW-FOAMING, MEDICAL GRADE DETERGENT SOLUTION, THEN FOLLOWING THE CLEANING AND STERILIZATION PROCEDURES GIVEN IN THIS CHAPTER. FAILURE TO REPROCESS THE INSTRUMENT IMMEDIATELY AFTER USE, OR USING OTHER THAN A MEDICAL-GRADE DETERGENT MAY CAUSE CORROSION AT THE METAL PARTS LIKE THE GRASPING JAWS. THIS COULD IMPAIR THE OPERATION OF THE INSTRUMENT OR CAUSE A PART OF IT TO BREAK AND/OR FALL OFF INSIDE THE PATIENT. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
Additional Manufacturer Narrative
THE DEVICE EVALUATION IS ONGOING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GRASPING FORCEPS EXHIBITED A BROKEN PART. NO REPORTS OF PARTS FALLEN INTO BODY. NO REPORTS OF PATIENT INVOLVEMENT.