MalfunctionFLE
AMSCO 600 STERILIZER
Received Jun 13, 2024 · Event occurred May 15, 2024
Report 3005899764-2024-00026 · MDR key 19534380
Device
Generic name
Sterilizer
Manufacturer
Steris Mexico, S. De R.l. De C.v.Product problems
- Leak/Splash
- Leak/Splash
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE USER FACILITY REPORTED THAT WATER WAS LEAKING FROM THEIR AMSCO 600 STERILIZER. NO REPORT OF INJURY.
Additional Manufacturer Narrative
A STERIS SERVICE TECHNICIAN ARRIVED ONSITE FOLLOWING THE REPORTED EVENT TO INSPECT THE STERILIZER AND FOUND THE MAIN WATER FEED LINE TO THE VACUUM PUMP SEPARATED AT THE CRIMP FITTING. AFTER FURTHER INSPECTION, THE TECHNICIAN FOUND THE CRIMP WAS DAMAGED. THE TECHNICIAN REPLACED THE HOSE, TESTED THE UNIT, CONFIRMED IT TO BE OPERATING ACCORDING TO SPECIFICATION, AND RETURNED IT TO SERVICE. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.