SPECTRA OPTIA®
Received Jun 13, 2024 · Event occurred Feb 1, 2024
Report 19528663 · MDR key 19528663
Device
Generic name
Separator, Automated, Blood Cell, Diagnostic
Manufacturer
Terumo Bct, Inc.Model number
03332619Catalog number
12220
Lot number
2310233141
Patient
56 YR · Female
- Insufficient Information
- Insufficient Information
Narrative
Description of Event or Problem
PT AT INFUSION CENTER FOR PLASMA EXCHANGE. THE FOLLOWING ALARMS OCCURRED ON THE OPTIA APHERESIS MACHINE DURING THE PROCEDURE. "LOW-LEVEL RESERVOIR DETECTED EXCESS FLUID" AND "PATIENT FLUID BALANCE MAY BE 5% HIGHER THAN REPORTED" SCREEN PROMPTS AND TROUBLESHOOTING WAS FOLLOWED AND PROCEDURE WAS COMPLETED WITHOUT ANY PATIENT COMPLICATIONS. THE LOW-LEVEL RESERVOIR ALARMS OCCURRED MORE THAN ONCE. NO PROBLEMS WERE SEEN WITH THE TUBING SET/FLUIDS OR PATIENT LINES. THESE SAME 2 ALARMS HAD ALSO OCCURRED ON A DIFFERENT PATIENT PLASMA EXCHANGE PROCEDURE THE DAY PRIOR WITH THE 2ND OPTIA MACHINE FOR WHICH THE SAME LOT# OF EXCHANGE SET WAS USED. EXCHANGE SET LOT#2310233141. TERUMO BCT WAS CALLED TO REPORT THE ALARMS. MANUFACTURER REPORTED THERE HAVE BEEN OTHER CALLS FOR THE SAME ALARMS. DETERMINED THE CAUSE WAS LIKELY THE EXCHANGE SET/TUBING. THE 9 REMAINING SETS FROM THIS LOT# WERE REMOVED FROM USE. EMAIL REQUEST SUBMITTED TO TERUMO BCT FOR CREDIT OR REPLACEMENT OF SETS. MANUFACTURER RESPONSE FOR TERUMO BCT EXCHANGE SET, (BRAND NOT PROVIDED) (PER SITE REPORTER). MANUFACTURER REPORTED THERE HAVE BEEN OTHER CALLS FOR THE SAME ALARMS. DETERMINED THE CAUSE WAS LIKELY THE EXCHANGE SET/TUBING. THE 9 REMAINING SETS FROM THIS LOT# WERE REMOVED FROM USE. EMAIL REQUEST SUBMITTED TO TERUMO BCT FOR CREDIT OR REPLACEMENT OF SETS.